ECG | Heart (Pulse) Rate | |||
Electrode | 3/ 5/12 Leads | Source | User Selectable : Smart, ECG | |
Numeric Data Display | 6 lead ECG with HR | Plenth , Nibp | ||
Heart Rate,VPC rate (per min.), | Range | Adult 15-300 bpm | ||
ST Level, respiration rate, | Pediatric 15 - 350 bpm | |||
Pulse rate sync mark, | Accuracy | + 1 bpm or + 1 % | ||
respiratory sync mark | ||||
NIBP (Systolic,Diastolic,MAP), | IBP (Optional) | |||
SpO2 (Nellcor), Pulse rate, | Channels | 2 Nos | ||
Temperature, Heart rate sync m | Pressure | ART , PA , CVP , RAP , LAP | ||
Lead | I,II,III,aVR, aVL, aVF, V (5 Lead) | ICP , P1 , P2 | ||
I,II or III (3 Lead) | Range | 0 - 300 , 6 - 120 , 10 - 300 mmHg | ||
I,II,III,aVR, aVL, aVF, V1-V6 (12 Lead) | Accuracy | + 1 mmHg or + 2% | ||
Gain Selection | x1/4, x1/2, x1 and x2 | |||
Frequency Response | Diagnostic : 0.05 to 90 Hz | Environment | ||
Monitor : 0.5 to 35 Hz | Operating Temperature | (10 - 40 C) | ||
Surgery : 1-15 Hz | Storage Temperature | (-5 - 50 C) | ||
Calibration Signal | 1(mV p-p), Accuracy: +5% | Humidity | 15% to 90% | |
Analysis | ST Segment & Arrhythmia Detection | |||
Display Channel | 6 Channels | Trends | ||
Sweep Speed | 6.25,12.5,25 or 50 mm/s | Memory | 200 Hr at 30 seconds intervals for | |
a NIBP reading every 5 minutes | ||||
Pulse Oximetry (Nellcor) | Display | Tabular, graphical | ||
Range | 0 to 100% | |||
Resolution | 1% | General | ||
Accuracy | 70% to 99% range +1% | Dimensions | 320 (L) x 280 (B) x 135 (T) mm | |
40 to 69% range + 1% | Weight | 3.5 Kg | ||
Method | Dual wavelength LED | Power | AC 100-240 V , 50/60 Hz | |
Power Consumption | 35 W | |||
Nibp | Battery | 14.8 V / 2200 Mah Li-Ion | ||
Technique | Oscillometric during inflation | Recharge Time | 4.5 hours | |
Range | Adult : 0-300 mm Hg | Operating Time | > 2 Hours typical | |
Pediatric: 0 - 250 mm Hg | Alarm Silence | 120 seconds | ||
Neonate : 0-135 mm Hg | Alarm Type | 3 Level adjustable Audio-visual alarms | ||
Measurement Cycle | <40 sec, typical | for each parameter, Sensor/Cuff | ||
Cycles (Selectable) | 2,3,5,10,15,30 min1,2,4 | disconnect, Low battery | ||
Max. allowable cuff pressure | Adult : 297 +3 mm Hg | Electrical Safety | ESU , Defibrillator protection | |
Pediatric : 240 +3 mm Hg | Connectivity | WiFi/ USB/ Ethernet port/ LAN | ||
Neonate : 147 +3 mm Hg | ||||
Resolution | 1 mm Hg | EtCo 2 (Optional) | ||
Transducer Accuracy | + 2 mm Hg or +2% of reading | Transducer Type | Side stream CO2 sensor | |
Modes | Auto, Manual & STAT | Range | 0-150 mm Hg | |
Resolution | 0.1 mm Hg | |||
Temperature ( Dual Channel) | Accuracy | 0-40mmHg +/-2mm | ||
Range, Accuracy | 0 - 50 C ; + 1% | 41-70 mmHg +/-5% of reading | ||
Display Resolution | +0.1 C | 71-100 mmHg +/-8% of reading | ||
Probe | YSI 400 AND YSI 700 Series | 101-150 mmHg +/-10% of reading | ||
Probe Gradation | 1/10 Degree | |||
Respiration Rate | ||||
Method | Impedance Pneumography | |||
Rate | Adult : 0 ~ 120 rpm | |||
Neo / Ped 0~ 150 rpm | ||||
Threshold Control | Yes | |||
Resolution | 1 rpm | |||
Accuracy | + 2 breath / min |